Products
EX-TEC® HS 680
EX-TEC® HS 680
The new standard with combination measuring instruments for gas utilities
A new milestone in gas measuring technology is marked by the
EX-TEC® HS 680.
The high-end model of the new range of instruments combines ergonomic handling and technology in one completely redesigned instrument concept. It complies with the requirements of the future appendix to G 465-4.
At the same time the measuring instrument weighs 30 % less than its predecessor EX-TEC® SR 6.
Features
• Innovative operating concept with jog dial, „soft keys" and menu guidance
• Large matrix display with background illumination
• Selective infrared sensors for methane and carbon dioxide
• Calibration on methane (standard), propane and butane (optional)
• Fast and highly sensitive semi-conductors
• Three optional electrochemical sensors for oxygen and toxic gases
• Ethane analysis to determine natural gas/marsh gas without further accessories (optional)
• Quick charging in 4 h, regular charging 10 h, even without charging station
• Power supply via 4 disposable Mignon accumulators or primary cells
• EX-protection according to ATEX 100a IIB, IIC with leather bag
• PC communication via USB
• Combined carrying and display bow
Applications / measuring ranges / sensors
|
Application |
Measuring range |
Sensors |
|
Inspection above ground |
1 ppm – 10 vol.% CH4 |
Gas sensitive semi conductor, |
|
Measuring in |
0,0 – 100 vol.% CH4 |
Infrared sensor |
|
Inspection of |
1 ppm – 100 vol.% CH4 |
Gas sensitive semi conductor, |
|
Inspection indoors |
1 ppm – 100 vol.% CH4 |
Gas sensitive semi conductor, |
|
Warning ExTox |
0 – 100 % LEL CH4 |
Infrared sensor |
|
0 – 5 vol.% CO2 |
Infrared sensor |
|
|
0 – 25 vol.% O2 (optional) |
Electrochemical |
|
|
0 – 100 ppm H2S (optional) |
Electrochemical |
|
|
0 – 500 ppm CO (optional) |
Electrochemical |
|
|
Measurement |
0,0 – 100 vol.% CH4 |
Infrared sensor |
|
Ethane analysis |
CH, CH4, C2H6, C3H8 |
Gas chromatograph |
Technical data
|
Operating time: |
min. 8,0 h |
|
|
Power supply: |
4 NiMH accumulators, rechargeable or |
|
|
Protection class: |
IP 54 |
|
|
Operating temperature: |
-10 °C – +40 °C |
|
|
Storage temperature: |
-25 °C – +50 °C |
|
|
Pressure: |
950 hPa – 1100 hPa |
|
|
Humidity: |
15% rF – 90% rF |
|
|
Dimensions: |
(W x H x D) 148 x 57 x 205 mm |
|
|
Weight: |
approx. 1000 g |
Accessories
|
|
|
|
|
|
|
|
|
|
|
- Water Leak Detection Systems
- Pipe and Sewer Spring Machines
- Gas Leak Detection Systems
- High Pressure Jetting Machines
- Ultrasonic Measuring Systems
- Electromagnetic Measure Systems
- Pipe & Cable Locators
- Fiber Optic Test and Maintanance Equipments
- Cable Fault Locators
- Magnetic Locators
- Holiday Detectors
- Pipeline Products
- Trenchless Products
- Inspection Products
- Pipe Plug